> Understanding Calibration Statuses

Datastream 7i assigns calibration statuses at the calibration header level and at the test point level. Calibration statuses are assigned based on the data entered for test points on the Calibration tab of the Work orders form (WMJOBS).

The Calibration status is based on the status of all its test points. The calibration fails if one test point fails. The Calibration status passes only when all of the test points pass. Test points are not considered passing if the test point status is set to recalibrate. For all other cases, the Calibration status is incomplete.

Datastream 7i determines the status of an individual test point based on the based on the Device reading for the test point and the specified tolerance for the device. There is an "as found" status and an "as left" status for each test point. The as found status indicates the status of the test point prior to the calibration process. The "as left" status indicates the status of the test point after the calibration test point.

Datastream 7i determines the status of the calibration header whenever a test point is inserted, updated, or deleted. If any one of the test points fails, then the calibration header also fails. A failing test point must be recalibrated until it passes.

The following diagram illustrates how upper and lower limits and tolerance ranges relate to a test point and its status.

There are a number of measurements and standards that can be set for test points on a category and/or an equipment level.

Refer to the following list for an explanation of the different calibration statuses.

Refer to the following example for further clarification of calibration statuses.

Example